Ptychographic X-ray computed tomography at a high-brilliance X-ray source

DOI: 10.1364/OE.27.000533 DOI Help

Authors: Simone Sala (University College London; Diamond Light Source; University of Southampton) , Darren Batey (Diamond Light Source) , Anupama Prakash (National University of Singapore) , Sharif Ahmed (Diamond Light Source) , Christoph Rau (Diamond Light Source) , Pierre Thibault (University of Southampton)
Co-authored by industrial partner: No

Type: Journal Paper
Journal: Optics Express , VOL 27

State: Published (Approved)
Published: January 2019
Diamond Proposal Number(s): 16199

Open Access Open Access

Abstract: Ptychographic X-ray computed tomography is a phase-contrast imaging technique capable of retrieving three-dimensional maps of the index of refraction of the imaged volumes with nanometric resolution. Despite its unmatched reach, its application remains prerogative of a limited number of laboratories at synchrotron sources. We present a detailed description of an experimental procedure and a data analysis pipeline which can be both exploited for ptychographic X-ray computed tomography experiments at any high-brilliance X-ray source. These have been validated at the I13-1 Coherence Branchline within the first experiment of its kind to be successfully carried out on a biological sample at Diamond Light Source.

Subject Areas: Technique Development


Instruments: I13-1-Coherence

Documents:
oe-27-2-533.pdf