Publication

Article Metrics

Citations


Online attention

Multi-scale multi-dimensional imaging at I13-coherence branchline in Diamond Light Source

DOI: 10.1063/1.5084623 DOI Help

Authors: S. Cipiccia (Diamond Light Source) , D. Batey (Diamond Light Source) , X. Shi (Diamond Light Source) , S. Williams (Diamond Light Source) , K. Wanelik (Diamond Light Source) , A. Wilson (Diamond Light Source) , P. Martin (University of Bristol) , T. Scott (University of Bristol) , C. Rau (Diamond Light Source)
Co-authored by industrial partner: No

Type: Conference Paper
Conference: 13th International Conference on Synchrotron Radiation Instrumentation – SRI2018
Peer Reviewed: No

State: Published (Approved)
Published: January 2019

Open Access Open Access

Abstract: The Coherence branch of the I13 beamline in Diamond Light Source is specialized in reciprocal space imaging using Ptychography and Bragg CDT techniques. The experimental hutch is 220 m from the synchrotron source, providing the x-ray beam with intrinsic high spatial coherence. The capacious hutch allows variable sample-detector distance up to l4.5 m. This, in combination with the use of the large area photon counting detector Excalibur, enables fields of view as large as 500 μm, resolution down to tens of nanometers and acquisition rate in excess of l00 Hz. Recently the ptychography scan has been combined to the x-ray fluorescence scan to simultaneously access the chemical and structural information of the sample.

Journal Keywords: Synchrotrons; Ptychography; Coherence; Crystal lattices; X-ray fluorescence; Optical metrology

Subject Areas: Technique Development, Physics


Instruments: I13-1-Coherence

Documents:
4545hth.pdf