Publication

Article Metrics

Citations


Online attention

Design of powder diffraction beamline (BL-I11) at Diamond

DOI: 10.1524/zksu.2007.2007.suppl_26.153 DOI Help

Authors: Chiu Tang (Diamond Light Source) , Stephen Thompson (Diamond Light Source) , Tim Hill (Diamond Light Source) , G Wilkin (Diamond Light Source) , Ulrich Wagner (Diamond Light Source)
Co-authored by industrial partner: No

Type: Journal Paper
Journal: Zeitschrift Fur Kristallographie , VOL 2007 , PAGES 153 - 158

State: Published (Approved)
Published: November 2007

Abstract: We present the design of a synchrotron x-ray beamline (BL-I11) dedicated to powder diffraction experiments, currently under construction at Diamond Light Source (DLS). An in-vaCLIurn Lindulator will provide an intense and highly collimated x-ray Source and with the necessary optics, a high purity beam of low energy-bandpass x-rays in the range 5-30 keV will be delivered at the sample. The diffraction instrument is designed to have the flexibility to house a variety of sample environments and to have two detection systems to collect high quality diffraction data, i.e. multi-analysing crystals for high angular resolution experiments and a fast position sensitive detector for tirne-resolved studies. BL-I11 will be a powerful state-of-the-art dedicated user facility for high resolution (d-space and time) powder diffraction studies of polycrystalline materials under ambient and non-ambient conditions.

Journal Keywords: Diamond Light Source; Synchrotron X-Rays; Powder Diffraction

Subject Areas: Technique Development


Technical Areas: